Mesis PressoMedical 3.0 ZAM-03 Manual De Usuario página 50

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 105
Table 6 – Input a.c power PORT
Phenomenon
Electrical fast
transients/bursts a
Surges a) b) j) o)
Line – to - line
Surges a) b) j) k) o)
Line - to - ground
Conducted
disturbances
induced by RF fields
c) d) o)
Voltage dips f ) p) r)
Voltage
interruptions f ) i)
o) r)
a) The test may be performed at any one power input voltage within the ME EQUIPMENT or ME SYSTEM
RATED voltage range. If the ME EQUIPMENT or ME SYSTEM is tested at one power input voltage, it is
not necessary to re-test at additional voltages.
b) All ME EQUIPMENT and ME SYSTEM cables are attached during the test.
c) Calibration for current injection clamps shall be performed in a 150Ω system.
d) If the frequency stepping skips over an ISM or amateur radio band. This applies to each ISM and
amateur radio band within the specified frequency range.
Basic EMC standard
or test method
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
IMMUNITY TEST LEVELS
Professional healthcare facility environment
Not applicable
±2kV
100kHZ repetition frequency
Not applicable
±0,5kV, ±1kV
Not applicable
±0,5kV, ±1kV, ±2kV
Not applicable
3V m)
0,15MHz – 80MHz
6V m) in ISM bands
between 0,15MHz and 80MHz n)
80% AM at 1kHz e)
Not applicable
0% UT; 1cycle and 70% UT; 25/30cycles h)
Single phase : at 0°
Not applicable
0% UT; 1cycle and 70% UT; 25/30cycles h)
Single phase : at 0°
Not applicable
0% UT; 250/300 cycle h)
50

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido