Chemical Method
DPD
Appendix
Interferences
Persistant Interferences
All oxidising agents in the samples react like chlorine, which leads to higher results.
Removeable Interferences
Interference from Copper and Iron (III) are eliminated by the addition of EDTA.
Concentrations above 4 mg/l Chlorine, in the event of using fluid reagents, can lead
to results within the measuring range of up to 0 mg/l. In this case, the sample must
be diluted with chlorine-free water. 10 ml of the diluted sample should be mixed with
the reagent and the measurement taken again (plausibility test).
Interference
CrO
2-
4
MnO
2
Conformity
EN ISO 7393-2
determination of free, combined and total
a)
EN   Method Reference Book   1.0
from / [mg/l]
0,01
0,01
Chlorine L / M101