9.3 Feature Details
AUTOMATIC ANTI-CRUSHING FUNCTION
The electric kit features an automatic patient anti-crushing function which blocks the descent
of the lifting arm in the event of accidental contact between the patient and an obstacle.
Image for illustrative purposes
operations in the event that the remote shows some problems. operation, in particular:
Button 3: 'up' button for lifting
Button 4: leg opening button
Button 5: button to close the legs
Button 6: 'up' button for lifting
There is a 3 LED indicator for the battery charge status. This indicator is located both on the
control unit (ref. 2 in the photo above) and on the remote control (ref. 6 in the following photo).
Three lit LEDs indicate a complete charge of the battery, while a single lit LED indicates that
the battery is low and we need to recharge it.
HOW TO CHARGE THE BATTERIES
•
Connect the adapter to the mains socket and then to the input of the control unit after
removing the relative protective cap
•
There is an orange LED on the charger adapter which is orange during the charging phase.
Charge the battery until the LED turns green
•
Before proceeding with the use of the lift remember to always disconnect the adapter to
recharge the batteries!
•
Charge the batteries for at least 6 hours before using the lift
9.4 Instructions for using the remote control
The remote control is connected to the control unit via an extendable spiral cable.
EMERGENCY STOP
On the control unit there is a red "STOP" button (ref. 1)
which, if pressed, immediately deactivates all commands.
To restore operation, turn the red knob clockwise until it
snaps up and returns to the rest position.
EMERGENCY COMMANDS
There are four buttons on the control unit (two for ascent
/ descent and two for opening / closing the legs) that can
be used instead of those on the remote control to perform
MECHANICAL LOWERING OF THE ACTUATOR
In case of use of the manual / mechanical descent, make
sure that it is carried out in conditions of total safety for the
patient. To activate the manual / mechanical descent, pull
the red lever in the lower part of the actuator upwards and
safely accompany the lowering of the device. Once lowering
is complete, return the red lever to the rest position.
INDICATION OF THE BATTERY STATE OF CHARGE