Technical data
Pump unit
Vacuum pump type
Vacuum range at 0–2000 MAMSL
Weight
Battery capacity
Dimensions
Protection class
Applied part
BLE radio frequency band
IP protection class
Life time of device incl. power adapter
Operating conditions
Storage and transport conditions
14
moderate vacuum, low flow rate
30–330 mbar
620 g
2600 mAh
210 × 150 × 90 mm
Class II
Class BF
2.4 GHz ISM
IP 22 (protected against solid objects up to 12 mm and protection
against dripping water when the housing is tilted at up to 15°)
800 hours
+5°C to +40°C
10% to 93% RH
700–1060 hPa
-10°C to +60°C
10% to 93% RH
700–1060 hPa