Zeiss Definite Focus.2 Manual Breve página 14

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 45
ZEISS
1
CAN port
2
Control 1 port
3
Control 2 port
4
Detection port
5
Service port
6
Power connector
7
Power switch
Fig. 2
Focus Controller.2, rear view
1
On/Off push-button
2
Display
Fig. 3
Focus Controller.2, front view
12
SWITCHING ON SAFELY
10
The Axio Observer microscope is delivered in a
partially assembled condition. It can be set up and
calibrated by the customer himself or through the
ZEISS service. Please refer to chapter INSTALLATION
of the complete Axio Observer and Definite Focus 2
operating manuals.
The following switching on sequence must be
followed strictly:
• To switch on Definite Focus.2 initially switch on
the power switch (Fig. 2/7) on the rear of Focus
Controller.2.
• Then start Definite Focus.2 via the On/Off push-
button (Fig. 3/1) on the front panel of the
controller. A beep will indicate that the system
has been switched on. While Definite Focus.2 is
booting, the blue LED of the on/off push-button
flashes and the current state of the device
appears on the display.
• Switch on the Axio Observer microscope after
starting Definite Focus.2 (see also the complete
operating manual of the microscope).
Displays during the boot routine on the display of
the Focus Controller.2 (Fig. 3/2):
424533-6044-008
SWITCHING ON SAFELY
"System Starting" - Initializing Definite
Focus.2.
"Detecting Stand" - Connecting to the
microscope. If the microscope is not ready,
the start routine of Definite Focus.2 will stop
here and wait. Once the microscope is ready,
the connection is established and the display
will show "Standby".
"Standby" - Definite Focus.2 is in standby,
equivalent to the Off position on the
touchscreen of the microscope. The device is
now ready to be used.
Definite Focus.2
09/2017