5 Mounting / exchanging the saw blade

If applicable use thick gloves to prevent injury when handling the saw
Only ever connect the HEBU oscillating cast cutting saw to the mains

5.1 Releasing the retaining screw / removing the old saw blade

1. Place the cast cutting saw on a stable underlay (table).
2. Apply one fork wrench at the drive shaft behind the saw blade,
and the other one at the front at the retaining screw (see fig. 1).
3. Hold the rear wrench firmly and use it to fix the cast cutting saw.
4. Release the retaining screw using the front wrench.
5. Remove the retaining screw and the fixing ring.
Ensure that you do not lose any components.
6. If applicable, remove the old saw blade.
Danger of injury due to unintentional activation.
Pull out the plug before changing the saw blade.
blade.
when it is switched off.
Fig. 1:
Releasing the retaining screw
EN
23