Problem
No aerosol is formed
The Valve Balloon does
not fill up.
The device switches off
(automatic switch-off).
Plant particles in the Val-
ve Balloon or Tube (when
using cannabis flos).
10. H O T A I R G E N E R AT O R
Possible cause/remedy
Check whether the set and actual temperatures are
the same (at least 180°C (356°F)).
Check whether the Filling Chamber and/or Valve
Balloon or Tube are correctly attached.
Check whether dronabinol has dropped onto the Drip
Pad or whether there is fresh, unused cannabis floss
in the Filling Chamber.
Check whether the pump is turned on.
Check whether the Filling Chamber and/or Valve
Balloon are correctly attached.
The automatic switch off has intervened. Switch the
heater or pump on again.
Check whether the Screen in the Filling Chamber Cap
is clogged or is not inserted into the notch properly.
Check whether the Filling Chamber was cleaned
according to the Instructions for Use.
122