MAINTENANCE
Maintenance
The following pages provide instructions for maintenance, cleaning, calibrating, and
troubleshooting of the scale. Maintenance operations other than those described in this manual
should be performed by qualified service personnel.
Caution: Before first use, or after long periods of non-use, check the scale for proper
operation and function. If the scale does not operate correctly, refer to qualified service
personnel.
1. Check overall appearance of the total scale for any obvious damage, wear, and tear.
2. Inspect the AC adapter cord for cracking or fraying, or for broken/bent prongs.
Note: This scale is an extremely sensitive weighing device. Items such as clothing or towels
placed or dropped on the scale may cause a scale to activate. In order to prevent battery
depletion, do not store or leave anything on the scale.
Cleaning and Disinfecting
Proper care and cleaning is essential to ensure a long life of accurate and effective weighing.
®
1. Health o meter
Professional recommends using one of the following solutions on a soft
cloth or disposable wipe:
● mild soap and water solution
● 70% isopropyl alcohol
● solution with 1-5% hydrogen peroxide concentration
After cleaning/disinfecting, wipe with a cloth dampened with water and then with a clean dry
cloth. To avoid residue build-up or damp surfaces, ensure all scale parts are completely dry
after cleaning.
2. Never use rough or abrasive materials to clean the scale, as these will damage the scale's
finish.
3. Do not submerge the scale in water or any other liquid.
4. Do not pour or spray fluids directly on the scale.
Replacement Parts
Parts may be available if a piece of the scale needs to be replaced. Contact Health o meter
Professional Scales Customer Service at 1-800-815-6615 to inquire about the availability of these
replacement parts.
Part #
499KLROD
499KLWHEELS
499KLFOOT
499KLMAT
BATTDOOR500
499KLPILLARFOOT
ADPT30
P/N UM499KL 499KG Rev 20201201_EngOnly
Description
Optional Height rod
Set of 2 wheels
Foot for base
Replacement mat for platform
Battery compartment door
Foot for pillar
Power adapter
22