Kipp & Zonen SMP12 Instrucciones De Servicio página 17

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 140
Fig. 1: Connection to RS-485 converter
i
In the pyranometer, the blue RS-485 ground is connected to the brown and black earth, and the white power
supply line is connected to the red power supply line.
4 Make sure that the power supply is switched off.
4 Connect the red to the white cable and the brown to the black cable on the power supply unit.
4 Connect the orange, gray and blue cables to the RS-485 converter.
4 Align the indentation on the connector with the indentation on the pyranometer's connection socket.
4 Plug the connector into the connection socket.
4 Turn the locking ring clockwise and tighten it hand tight to secure the connector.
NOTE! The seal may be damaged by overtightening!
4 Switch on the power supply.
4 Switch on the computer.
i
It may take three hours for the pyranometer to reach a stable temperature. During this time, the irradiation
measurements may deviate from the final measurements.
SMP12 Smart Pyranometer | EN
Red – power supply +10 up to 30 VDC
White – power supply +10 up to 30 VDC
Green – not connected
Orange – RS-485 data +
Gray – RS-485 data +
Blue – RS-485 ground
Brown – earth
Black – earth
Installation and commissioning | 4
RS-485 USB converter
17

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido