13.
Note on electromagnetic compatibility
Guidelines and manufacturer's declaration - electromagnetic immunity
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short inter-
ruptions
and voltage variations in power
supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
Conduced RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
NOTE: UT is the AC mains voltage before applying the test level.
IEC 60601-1-2 Test level
± 8 kV
for contact discharge
± 2, ± 4, ± 8, ± 15 kV
for air discharge
± 2 kV
power supply lines
± 1 kV
input/output lines
100 kHz repetition frequency
± 1 kV
line(s) to line(s)
± 2 kV
line(s) to earth
0% 0.5 cycle
At 0º, 45 º,
90 º, 135 º, 180 º, 225 º,270 º
and 315 º 0% 1 cycle
And 70% 25/30 cycles
Single phase: at 0 0% 250 cycle
30 A/m
50 Hz/60 Hz
150KHz to 80MHz:
3Vrms 6Vrms (in ISM and
amateur radio bands)
80% Am at 1kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Compliance level
± 8 kV
for contact discharge
± 2, ± 4, ± 8, ± 15 kV
for air discharge
N/A
N/A
N/A
30 A/m
50 Hz/60 Hz
N/A
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz