Hand Check Guide
Use this guide to adjust your cushion during Optional Setup.
A hand check is also recommended after performing Smart Setup so you will have a reference point if you need to adjust your cushion without
Smart Check (for example, if the device has been misplaced).
Before you perform a hand check, refer to "Important Safety Information" and read the warning about using an under-inflated cushion or an
over-inflated cushion.
1. Place your hand between yourself and the cushion. Feel for the
lowest bony prominence. You should be able to slightly move
your fingertips.
2. If it feels like the cushion has too much air or not enough air,
adjust as needed and see "Check Instructions" and "Overview."
Hard Reset Instructions
Before you perform a new Setup, you must clear the setting on Smart Check by performing a Hard Reset.
Note: Smart Check may be attached or detached when performing the Hard Reset.
1. Make sure there are no steady or flashing lights. If there is a
steady or flashing light, wait until it goes out before proceeding.
2. Simultaneously press and hold both the Check Button and the
Setup Button for five (5) seconds. Each of the lights will flash one
time, indicating that your previous setting has been removed.
3. After the Hard Reset is complete, follow all setup instructions as
indicated in this manual. (See "Smart Setup" or "Optional Setup.")
Troubleshooting
For additional assistance, contact the equipment provider, distributor, or Customer Support.
Problem
No lights are displayed on Smart Check after you
press the Check Button.
Smart Setup: After you press the Check Button,
Smart Check still displays the red "add air"()
light, but you have confirmed that the cushion is
overinflated and that the quick disconnect is securely
attached.
Optional Setup: You do not see a green light after
pressing and holding the Setup Button for four (4)
seconds.
Assistance is needed to use or maintain Smart
Check; you encounter unexpected problems; or a
component is damaged.
Smart Check Operation Manual
Warning: Make sure that there is at least a ½ in. (1.5 cm)
of air supporting the individual's bony prominences.
Action
Make sure the pull tab has been removed from the battery compartment. Confirm the
battery type and battery orientation. Replace the batteries if needed. See "Maintenance."
Contact Customer Support if the issue continues.
Inspect the O-ring on the cushion's quick disconnect for wear or damage. If damage is
found, contact Customer Support.
Inspect the cushion for a slow leak. See "Troubleshooting" in the manual provided with the
cushion.
Repeat Part 2 of "Optional Setup Instructions." If you still do not see a green light, note
that Optional Setup may not be compatible with your physical needs or preferences. Refer
to the "Overview" and "Smart Setup Instructions," or contact your clinician or Customer
Support.
Contact Customer Support.
7
5s
+