Troubleshooting guide
Problem
No power "on" unit
when the power
switch is on.
No nebulization or
low nebulization rate
when the power is
"on".
The device is very
hot.
Cause
The AC power cord
is not plugged into an
electrical outlet.
No medication in the
medication tank. Too
much or too little medi-
cation in the medication
tank.
The nebulizer kit is not
assembled correctly.
The nebulizer kit is titled
at an incorrect angle.
The air tube is attached
incorrectly.
The air tube is folded or
damaged. The air tube
is blocked.
The compressor is
covered. The device
has been used longer
than 20 minutes.
35
Solution
Turn the power switch
off. Plug the power plug
into an electronical outlet.
Turn the device on.
Add the correct amount
of prescribed medication
to the medication tank.
Make sure the nebulizer
kit is assembled correctly
and the inhalation acces-
sory is attached correctly.
Hold the nebulizer kit
correctly. Do not tilt the
nebulizer kit so that the
angle of the kit is higher
than 45 degrees.
Make sure the air tube
is attached correctly to
the compressor and the
nebulizer kit.
Make sure the air tube
is not folded, kinked or
bent. Inspect the air tube
for any damage. Replace
the air tube if damaged.
Do not cover the com-
pressor with any type of
cover during use. Turn
the device off. Wait 40
minutes before using the
device again.