ADJUST THERMOCOUPLE OFFSET
1. Insert the thermocouple into a known temperature
(T) until the display equal to known temperature (T).
ex: ice point at 0°C / boiling water at 100°C
2. Press "p" or "q" to add or subtract the value.
3. It can be adjusted ±6°F(±3°C) of default.
If you can't adjust your T/C, please check your T/C or
send the meter to be calibrated.
4. Press "
SPECIFICATION
Electrical
Temperature
Scale
Measurement
Range
Accuracy
Temperature
Coefficient
" key to confirm.
Celsius or Fahrenheit user-selectable
K-Type -100°C to 1372°C,
(-150°F to 1999°F)
J-Type -100°C to 1200°C,
(-150°F to 1999°F)
Accuracy is specified for operating
temperatures over the range of 18°C
to 28°C (64°F to 82°F), for 1 year, not
including thermocouple error.
± (0.1%rdg+1°C) on °C
± (0.1%rdg+2°F) on °F
0.1 times the applicable accuracy
specification per °C from 0°C to 18°C
and 28°C to 50°C (32°F to 64°F and 82°F
to 122°F).
6