3. Remove the fuse by gently prying one end loose, and take out the fuse from its bracket.
4. Install ONLY replacement fuses with identical amperage and voltage specified: F1A240V
5. Rejoin the battery door and the case bottom, and reinstall the screws.
6. Do not operate the Meter with the case opened.
Replacement of the fuse is seldom required. Buring of a fuse always results from improper
operation.
Maximum Voltage applied to any Terminal
mA Input Terminal
Maximum Display
Operating Temperature
Storage Temperture
Relative Humidity
Auto Power Off
Accuracy: ±(a % of rdg + b digits), guarantee for 1 year.
Operating Temperature: 23°C±5°C
Relative Humidity: <75%
AC VOLTAGE MEASUREMENT
Range
6.000V
60.00V
600.0V
1000V
Remarks:
Imput Impedance: ≥10MΩ
Displays effective value of sine wave
Frequency Response: 50Hz~1kHz
42
Table 6. General Specifications
Altitude
Batteries
Low battery
Size
Weight
ELECTRICAL SPECIFICATIONS
Resolution
1mV
10mV
100mV
1V
SPECIFICATIONS
1A / 240V Fast Fuse, High breaking capacity Fuse
6000, Updates 5 times per second
0°C ~40°C (32°F ~104°F)
-10°C ~50°C (14°F ~122°F)
≤75% @ 0°C to below 31°C
≤50% @ 31°C to above 40°C
Operating: 2000 meter
Storage: 10000 meter
Six AA 1.5V batteries AM3/LR6 (standard capacity
The Meter will be automatically off when the Meter
is not in use for more than 15 minutes
97mm x 202mm x 46mm
Around 505 g (including batteries)
Accuracy
50Hz ~60Hz
± (1%+3)
± (2%+3)
1000V rms
around 2450mAh)
LCD display
Input Protection
60Hz ~ 1kHz
± (2%+3)
1000V
± (2%+6)

Publicidad