Safety instructions
Number
Warning / Information symbol
1
2
3
Warnung! Heiße Oberflächen! Verbrennungsge-
fahr! Caution! Hot surface!
Gefährliche UV-Strahlung! Nicht direkt in Ofen-
strahlung / Flamme schauen!
Caution! Emission of UV radiation!
Kurzschlussgefahr!
Bedienung mit Schmuck verboten!
Danger of short circuit!
Handling with jewels not allowed!
4
Achtung! Bei geöffneter Tür Lampen ausschalten!
Warning! Switch off lamps when door is opened!
2.3
Requirements for the operating personnel
2.4
Safety instructions, transport and installation
24
The novAA 800 must only be operated by qualified specialist personnel instructed in
the use of the device. The instruction must also include conveying the content of this
short instruction and the user manual as well as the user manuals of other system
components.
In addition to the safety at work instructions in this user manual the generally applica-
ble safety and accident prevention regulations of the respective country of operation
must be observed and adhered to. The operator must ascertain the latest version of
these regulations.
All manuals and instructions must be accessible to the operating and service personnel
at any time!
Observe the following notes:
The novAA 800 is always installed by the service department of Analytik Jena or by
specialist personnel that was authorized and trained by Analytik Jena. Independent
assembly and installation are not permitted. Incorrect installation can create seri-
ous hazards.
Meaning and scope of application
Hot surfaces! Risk of burns at the hot graphite fur-
nace and burner! (Position: at the height adjust-
ment)
Read the operating manual before commencing
work. (Position: at the mains switch on the right-
side of novAA 800)
Hot surfaces! Risk of burns at the hot graphite fur-
nace and burner!
Dangerous UV radiation! Do not look into the
lamp beam, into the flame or the graphite furnace
without UV protection goggles.
Short circuit warning only for novAA 800 D + G
Do not wear any metallic jewelry (especially
around the neck). Not observing this instruction
may cause a short circuit in the electrically heated
furnace. The jewelry may become very hot and
cause burns.
The ultraviolet light emitted by the (deuterium)
hollow cathode lamps can cause eye and skin
damage. Turn off the lamp current via ASpect LS
software before opening the lamp chamber.
novAA 800