Components
Depth Indicator/
Lancet Holder
Adjustable Tip
Clear Tip
Ejection System
Sliding Barrel
Release Button
Performing a blood test with your fingertip
Step 1
Firmly insert the test strip into the meter test port in
the direction of arrow on test strip (arrow up). Please
do not insert the glucose test strip upside down.
When you insert the glucose test strip into the
glucose test meter, the power automatically turns on
with the code and temperature. Make sure the code
number matches on the test strip bottle (e. g. CODE
125). adia automatically recognizes the test strip code
number and adjusts the meter accordingly.
A blinking test strip will appear at the top of the
screen, which indicates that the meter is ready for
testing.
Caution
1. If you do not see the code displayed, pull the glucose test strip out of the
port, and re-start the procedure from the beginning.
2. If you do not see the code displayed match the code printed on the vial,
please try it again with another test strip. If the mismatch persists, please
contact your local representative for assistance.
3. Avoid testing under direct sunlight, for a more accurate test result.
4. If you apply your blood sample before a blinking test strip on the screen,
Er5 message will appear on LCD. Please refer to Page 39.