Inconsistent or Unexpected Test Results
If you continue to get unexpected results, check your system with control
solution. If you experience symptoms that are not consistent with your
glucose results, review and follow all the instructions in this manual. Do not
ignore symptoms or make significant changes to your diabetes control
program by yourself without informing healthcare professional of your
symptoms and/or concerns.
Caution
1. Low Glucose Results: If your result is lower than 70 mg/dL, it may mean
hypoglycemia (low blood sugar). This may require immediate treatment,
according to your healthcare professional's recommendations. Although this
result could be due to a test error, it is safer to treat first and you may test
another.
2. High Glucose Results: if your test result is higher than 180 mg/dL, it may
mean hyperglycemia (high blood sugar). If you are uncertain about your test
results, consider re-testing. Your health care professional will help you to
decide how you need to act. If the meter displays "HI" all the time, please
re-check your blood sugar. If the "HI" continues, please consult with your
healthcare professional immediately.
9. Warranty
The warranty is based on the legal provisions (unless otherwise agreed).
If you have any questions or complaints, please contact your local distributor.
A return of your blood glucose meter needs to be confirmed in advance.
10. Specifications
Sample Type
Sample Volume
Test Range
Reading Time
Calibration
Hematocrit
Altitude
Operating Temperature
Operating Humidity
Test Strip Storage Temperature
Display Type
Dimension (HxDxT)
Weight
Power Source
Battery life
Capillary and/or venous whole blood
0.5 µl
20-600 mg/dl (1.1-33.3 mmol/L)
5 seconds
Plasma–equivalent
20-60 %
up to 3048 meter (10,000 feet)
10-40 °C (50-104 °F)
10-90 %
Store 2-30 °C (36-86 °F) and no direct
sunlight. Do not freeze.
LCD
81 X 52 X 16 ±1 (mm)
43±1g (Including Battery)
3V Li Battery (CR2032) X 1
One year after purchasing