Data averaging and
other signal
processing
Data Update Period
Operating
Environment
Storage & Transport
Environment
7 Clinical SpO2 Accuracy
The below table shows statistic distribution of an invasive
controlled desaturation study, which guided by ISO80601-2-61,
Annex EE, Guideline for evaluating and documenting SpO2
Accuracy in human subjects.
The statistic distribution displayed the accuracy distribution
between the range of 70% ~ 100%, which may helpful to user.
8s
1s
Temperature: +5°C to +40°C
Humidity:15% to 93%
non-condensing
Air Pressure: 70Kpa-106Kpa
Temperature: -25°C to +55°C
Humidity:15% to 93%
non-condensing
Air Pressure: 70Kpa-106Kpa
18