10 Manufacturer's EMC Declaration
1) All necessary instructions for maintaining BASIC SAFETY and
ESSENTIAL PERFORMANCE with regard to electromagnetic disturbances
for the expected service life.
2) Guidance and manufacturer's declaration - Electromagnetic Emissions and
Immunity.
Table 8 - Guidance and manufacturer's declaration - electromagnetic
emissions
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Table 9 - Guidance and manufacturer's declaration - Electromagnetic
Immunity
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601-1-2 Test level
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Power supply lines:±2 kV
input/output lines:±1 kV
line(s) to line(s): ±1 kV.
line(s) to earth: ±2 kV.
100 kHz repetition frequency
21
Compliance
Group 1
Class B
Not application
Not application
Compliance level
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15 kV air
Not application
Not application