Separation distances between RF communications equipment and
Platinum Series system
Recommended separation distances between portable and mobile RF
communications equipment and the Platinum Series system
The Platinum Series system is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of
the Platinum Series system can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Platinum Series system as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power
of transmitter (W)
0.01
0.1
1
10
100
Separation distance according to
frequency of transmitter (m)
d = 1.2√P
< 800 MHz
0.12
0.38
1.2
3.8
12
d = 2.3√P
≥ 800 MHz
0.23
0.73
2.3
7.3
23