Trouble removal
Problem
No Power
No Inflation
Err and stop working
Cuff leak
Please contact the distributor if you can't solve the problem, do not disassemble the unit by
yourself!
CAUSE
Monitor could not detect pulse
wave or cannot calculate the
blood pressure data
Measurement result is abnormal
(SYS≤35mmHg, DIA≤23mmHg)
Too loose cuff or air leakage
(Cannot inflate to 30mmHg within
15s)
The air tube is crimped
The sensor is sensing great
fluctuation in the pressure
The pressure that the sensor
sensing is over the limit
The demarcation is incorrect or
the device has not been
demarcated
Check
Whether battery has enough
power
Whether polarity is correct
Whether the plug is correctly
inserted
Whether the plug broken or leak
Whether move the arm when
inflate
Whether the chatting icon when
measured is on
Whether the cuff wrap too loose
Whether the cuff broken
CORRECTION
Check if the air releasing is too slow or
not. If it is too slow, please check if there
is any dust in the tube plug of the cuff
and the cuff port in the device. If yes,
please clean and start the measurement
again. If no, please send the device back
to the local distributor.
Occasionally-measure for one more time
/ Always - send it to local distributor
Tie the cuff correctly and make sure the
air plug is properly inserted in the unit
Correct it and make the measurement
again
Please keep quiet and don't move
Please send back to the local distributor
Please send back to the local distributor
Cause and solutions
Replace new one
Installation for proper placement
of the batteries polarities
Insert into the air socket tightly
Change a new cuff
Keep the body calm
Keep quite when measure
Wrap the cuff tightly
Change a new cuff
11