Electrical fast
transient/burst IEC
61000-4-4
Surge IEC 61000-
4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
Power frequency
(50Hz/60Hz)
magnetic field IEC
61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacture's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer
or the user of the device should assure that it is used in such an environment.
Immunity test
3 Vrms
150 kHz to 80 MHz 3
V RMS outside the
ISM band, 6 V RMS
in
Conducted RF
amateur bands
IEC 61000-4-6
80% AM at 1kHz
±2 kV for power
supply lines ±1 kV
for input/output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
0 % UT; 0.5 cycle at
0°,45°,90°, 135°,
180°, 225°, 270°,
315°
0 % UT ; 1 cycle
70 % UT; 25/30
cycle
0% UT; 250/300
cycle
30 A/m
50/60Hz
IEC 60601 test
level
the
ISM
and
Not applicable
Not applicable
Not applicable
30 A/m
50/60Hz
Compliance level
Not applicable
16
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the device requires continued
operation during power mains
interruptions, it is recommended
that the device be powered from
an uninterruptible power supply
or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Electromagnetic
environment - guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the device,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance