reading
2
The display will show "150" and "CO2 PPM" and start counting down in
seconds. The relevant CO
ventilation recommendation will be displayed in 2.5 minutes. Then it takes
around 30 minutes to have all the CO
If the AC power is interrupted, the monitor does not function; the display
and the backlight are turned off. The settings of your unit will be kept by the
internal memory chip so your monitor will function again after the AC power
resumes.
Minimum and maximum readings are automatically reset when the unit is
restarted.
Please note that common storage media such as memory chips, USB sticks
and external hard drives can suffer data loss due to data decay after more
than a year without power supply. We therefore recommend a regular
power supply to protect your settings.
Notes:
Make sure the place you put your monitor at is well ventilated (with a
CO
level near as low as 400 PPM possible) for at least an hour or so
2
everyday, or you move the monitor to such a well ventilated room for
an hour or so everyday to perform automatic baseline calibration.
Moving your monitor will affect the CO
minutes to have the reading stabilized again after you keep it in a fixed
position.
reading and the colour icon of air quality /
2
Calibration countdown (150 seconds)
reading stabilized.
2
reading. It takes up-to 5
2
EN3