The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an
environment.
Immunity test
Electrostatic discharge (ESD) IEC
61000-4-2
Electrical fast transient/burst IEC
61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and
voltage variations on power supply
input lines IEC 61000-4-11
Power frequency (50Hz/60Hz)
magnetic field IEC 61000-4-8
Note: Ur is the AC mains voltage prior to application of the test level.
36
Guidance and manufacturer's declaration - electromagnetic immunity
IEC 60601 test level
±8 kV contact
±15 kV air
±2 kV for power supply lines
± 1 kV line(s) to line(s)
<5% UT (>95% dip
in UT) for 0.5 cycle 40% UT (60% dip
in UT) for 5 cycles 70%UT(30%
dip in UT) for 25 cycles <5% UT (>95%
dip in UT) for 5 sec
l0V/m
Compliance level
±8 kV
Floors should be wood, concrete or ceramic tile. lf floor are
covered with synthetic materi- al, the relative humidity
contact ±15 kV air
should be at least 30%.
±2kV for power supply
Mains power quality should be that of a typical commercial
lines
or hospital environment.
±1 kV
Mains power quality should be that of a typical commercial
or hospital environment.
differential mode
<5% UT (>95%
Mains power quality should be that of a typical commercial
or hospital environment. lf the user of the DEVICE requires
dip in UT) for
continued operation during power mains interruptions,
0.5 cycle 40% UT (60%
it is recommended that the DEVICE be powered from an
dip in UT) for 5 cycles
uninterruptible power supply or a battery.
70% UT (30% dip in UT)
for25
cycles <5% UT (>950/odip
in UT) for 5 sec
l0V/m
Power frequency magnetic field should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
Electromagnetic environment
- guidance