LOW BATTERY INDICATION
Lo
The display will show
and need to be replaced.
AUTO SHUT-OFF
To extend battery life, the scale will automatically turn off
after approximately four minutes if no active weighing is
occurring. This feature is active with battery operation only.
CALIBRATION
For best results, calibrate the scale at regular intervals,
especially if using for prolonged periods. Calibration
weights are not provided with the scale.
CALIBRATION VALUES
MODEL #
H-2715
H-2716
H-2717
CAL E
The message
will
appear if the calibration
steps are not followed or
the wrong weight was used.
1.
Press and hold the UNIT
CAL button to start the
calibration process. The
CAL
display will show
NOTE: The calibration process can be aborted
by turning the scale off.
2. Press the ON/OFF ZERO
button to capture zero.
-C-
The display shows
while the scale stores
the zero load signal.
3. The display will show
C
xxx where xxx is the
calibration weight in
grams.
4. Place the appropriate calibration weight on the
platform.
PAGE 3 OF 9
OPERATION CONTINUED
when the batteries are weak
CALIBRATION POINT (g.)
200
1,000
2,000
.
5. Press the ON/OFF ZERO
button. The display shows
-C-
while the scale stores
the calibration point
signal.
6. After calibration, the
display returns to the
normal weighing mode.
1-800-295-5510
uline.com
0521 IH-2715