CONFIGURATION MENU
CALIBRATION
page 1 of 2
Operator must set the pump's maximum flow rate output. The units displayed in the Calibration
submenu are controlled by the units selected in the Units submenu.
IMPORTANT: The value entered (in Calibration) for the pump's maximum output is used to calculate the pump's
required output for the operating mode in Run Pump. The value is also used to calculate the pump's required run
speed in the PPM Feed submenu in Control Mode. For the most accurate dosing possible, determine the actual pump
output. Conduct a drawdown from a graduated calibration column against the discharge run and pressure, equivalent to
the intended installation.
NAVIGATION
Moves up in a menu, toggles
between options, or increase a value
Holding the
or
button causes the numbers to change rapidly.
Configuration
- Units
- Clock
- Calibration
S Series FW 4.01.02 US and Canada 800.683.2378, International 904.641.1666 www.stenner.com
continued
Moves down in a menu, toggles
between options, or decreases a value
Calibration
- GPD/LPD Calibration
- Signal Calibration
- GO TO CONFIG MENU
Calibration
- GPD/LPD Calibration
- Signal Calibration
- GO TO CONFIG MENU
Calibration
- GPD/LPD Calibration
- Signal Calibration
- GO TO CONFIG MENU
To Configuration menu
17
Moves one step back in
Sets a value in a menu
a menu, when permitted
GPD or LPD is selected in
Units under Configuration
Enter max pump flow
_ _ 0.0 GPD
adjust
set
Continued on next page
To
Calibration menu
When ENTER is
pressed for the
final digit in
Calibration,
the display
returns to the
Calibration menu