EN-24
4. MAINTENANCE
Cleaning
To ensure proper balance operation, keep the housing and platform clean. If necessary,
a cloth dampened with a mild detergent may be used. Check under the platform for
debris and remove. Keep calibration masses in a safe dry place. Unplug the AC
Adapter when not in use. For long term storage, remove the batteries.
Troubleshooting
SYMPTOM
No display.
Battery Indicator is flashing.
Incorrect weight reading.
Calibration procedures do
not work.
Unable to display weight in
a particular weighing unit.
Balance readings unstable.
Error code is displayed.
REMEDY
PROBABLE CAUSE
1. Connect AC Adapter.
1. Power Adapter not
connected.
2. Replace batteries.
2. Batteries are exhausted.
Replace batteries.
Batteries are weak.
1. Calibrate the balance.
1. Balance out of calibration.
2. Press the ON/ZERO Off
2. Balance was not
button with no weight on the
rezeroed before weighing.
platform, then weigh item.
1. Use correct masses.
Incorrect calibration
2. See error codes note ERR4.
masses being used.
1. Use Units menu to set
1. Weighing unit not
desired units ON.
activated in menu.
2. Exit Hold or Totalize mode.
2. Mode prevents unit
change.
Either move or shield
Balance location may
the balance from external
have drafts from air
air currents or vibration.
conditioning vents or
vibration from other
appliances nearby
affecting operation.
Review error code table
Various internal and
and take appropriate
external problems
action.
SCOUT PRO