Descargar Imprimir esta página

Medline MDS3002 Manual Del Usuario página 30

Monitor de presión arterial con manguito para brazo totalmente automático adulto grande esfigmomanómetro electrónico

Publicidad

Idiomas disponibles

Idiomas disponibles

For all ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The [MDS3002/MDS3002LA] is intended for use in the electromagnetic environment specified
below. The customer or the user of the [MDS3002/MDS3002LA] should assure that it is used in
such an environment.
IMMUNITY
IEC 60601test
test
Radiated RF
3 V/m 80 MHz
IEC 61000-
to 2,5 GHz
4-3
Compliance
level
level
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the
[MDS3002/MDS3002LA], including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of the
transmitter.
 
Recommended separation distance:
 
-30-
80 MHz to 800 MHz
800 MHz to 2,5 GHz

Publicidad

loading

Este manual también es adecuado para:

Mds3002la