portable and mobile RF communications equipment and the [MDS3002/MDS3002LA]
The [MDS3002/MDS3002LA] is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the [MDS3002/
MDS3002LA] can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the [MDS3002/
MDS3002LA] as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output
power of
transmitter
0,01
0,1
1
10
100
Recommended separation distances between
Separation distance according to frequency of transmitter
 
150 kHz to 80 MHz
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
 
0.12
0.38
1.2
3.8
12
-32-
 
800 MHz to 2,5 GHz
0.23
0.73
2.3
7.3
23