EN
maintenance / specifications
mAINTENANCE
All maintenance must be performed by a qualified DeVilbiss provider or authorized service
center.
� danger
Electric shock hazard . Do not remove compressor cabinet . All disassembly
and maintenance must be done by a qualified DeVilbiss provider .
SPECIFICATIoNS
Size ................................................. 7.5" W x 4.0" H x 7.2" D (19.1 cm x 10.2 cm x 18.3 cm)
Weight ............................................................................................................. 4.8 lbs. (2.18 kg)
Electrical Requirements
3655D ................................................................................ 2.5 amps, 115 VAC (±10%) 60 Hz
3655I.................................................................................. 1.4 amps, 230 VAC (±10%) 50 Hz
Sound Level .................................................................................................................... 58 dBA
Power Consumption
3655D ........................................................................................................ 140 watts maximum
3655I.......................................................................................................... 160 watts maximum
Maximum Compressor Pressure ................................................ 35 psig or greater (241 kPa)*
Compressor Free Air Flow ............................................................................8 lpm or greater*
Operating Temperature Range .............................................. +40° to +104°F (+5 to +40°C)
Operating Humidity ....................................................................... up to 95% non-condensing
Storage/Transport Temperature Range ................................. -40° to +158°F (-40° to +70°C)
Storage/Transport Humidity .......................................................... up to 95% non-condensing
Supplied Nebulizer Capacity ........................................................................................ 6 ml (cc)
Supplied Nebulizer Nebulization Rate ......................................................... 0.15 ml/min or greater*
Supplied Nebulizer MMAD ......................................................................... 5 microns or less**
*Conditions may vary based on altitude above sea level, barometric pressure, and temperature.
**Value determined with minimum flow rate through the nebulizer (20°C, sea level).
There are no hazardous materials used in the construction of this unit, therefore, there are
no known risks associated with disposal of the equipment.
This product (model 3655D) complies with the following electromagnetic compatibility standard: IEC
60601-1-2:2nd Edition
3655D Equipment Classifications:
Protection against electric shock:....................................................................................Class II
Degree of protection against electric shock: ................................................................ Type BF
Equipment not suitable for use in the presence of flammable anaesthetic mixture with air or
with oxygen or nitrous oxide
Mode of operation: ................................................................................. Continuous operation
A-655
9