1. ATTAChINg AND REmOvINg FILE
(1) Attaching
•
Press the push button to open the chuck. (Fig.1)
•
Insert the instrument handle into the chuck (Fig.2) and release the
button.
Caution
•
Make sure the file is mounted
firmly by pulling before use.
(Fig.3)
(2) Removing
•
Press the push button to remove the file.
2. SpEED
Allowable Max. Motor Speed............10,000 min
3. LUBRICATION
Lubricate with PANA-SPRAY Plus after each use and/or before autoclaving.
1. Push F-type spray nozzle attachment over the PANA-SPRAY Plus nozzle
until it seats firmly.
2. Shake the can 3-4 times to mix lubricant and propellant well.
3. Insert the F-type spray nozzle into the rear of the handpiece and spray for
approximately 2-3 seconds until the oil comes out of the handpiece head
or nose.
F-type Spray Nozzle
PANA-SPRAY Plus
Cautions on Lubrication
•
Be sure to hold the handpiece firmly to prevent the handpiece from
slipping out during the lubrication.
•
Keep the PANA-SPRAY Plus can upright.