4. CLEANINg AND STERILIzATION
Sterilization by autoclaving is recommended.
Sterilization required after each patient use as noted below.
Autoclaving procedure
1. Brush off dirt and wipe clean with alcohol-soaked cloth or paper
towel. Do not use a wire brush.
2. Remove reduction sheath from the motor and lubricate. Refer to
Section 3. Lubrication.
3. Insert into a sterilization bag and seal it.
4. Autoclavable up to a max. 135˚C. ex.) Autoclave for 20 min. at 121˚C,
or 15 min. at 132˚C.
5. Keep the handpiece in the autoclave pouch to keep it clean until
next use.
* Sterilization at 121˚C for more than 15 minutes is recommended by
EN13060 or EN ISO17665-1.
* Performing sterilization according to these instructions has minimal
effect on the instrument. Life span is generally determined by wear and
damage due to use.
Cautions on Sterilization
Skip dry cycle.
Heating the unit at the bottom of the chamber, may exceed the set value
of the unit.
Place the handpiece on the central or upper tray.
Do not autoclave the handpiece, when it is bagged together with other
instruments. This helps avoid discoloration and/or damage the plating of
the handpiece.
5