For consigned systems: DJO, LLC will repair or replace all or part of the 30BI-S and
30BI-SB systems for material or workmanship defects for the life of the customer
contract . Warranty and maintenance terms are specified in each consignment
contract .
For purchased systems: DJO, LLC will repair or replace all or part of the 30BI-S or
30BI-SB systems for material or workmanship defects for 3 years from the date of
sale .
For disposable cuffs, tubing and battery pack purchases: DJO, LLC will repair
or replace all or part of the cuffs, tube units and battery packs for material or
workmanship defects, not including normal depreciation of battery life or charge
capacity, for a period of six months from the date of sale .
Optional Tube Alarm Verification Test
The Tube Alarm Verification Test is provided for facility verification purposes only and
is not required prior to use .
Attach tube assembly to pump and connect power cord to electrical outlet . Simulate
a kinked tube situation by folding the tube over onto itself once and holding it for 4-6
minutes . The second cycle after the simulated kink should activate an audible alarm
and this symbol
will appear with "CHECK TUBES" on the graphical display and the
pump indicator lights on the side of the system will flash red .