Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The VenaFlow Elite is intended for use in the electromagnetic environment specified below .
The customer or the user of the VenaFlow Elite should assure that it is used in such an
environment .
Immunity Test
Electrostatic
Discharge
±6kV contact
(ESD)
±8kV air
IEC 61000-4-2
±2kV for power
Electrical Fast
supply lines
Transient/Burst
±1kV for input/
IEC 61000-4-4
output lines
±1kV differential
Surge
mode
IEC 61000-4-5
±2kV common
mode
Voltage
<5% U
dips, short
U
interruptions
and voltage
40% U
variations on
U
power supply
input lines
70% U
U
IEC 61000-4-11
<5% U
U
Power
Frequency
(50/60Hz)
Magnetic
3 A/m
Fields
IEC 61000-4-8
NOTE: U
is the a . c mains voltage prior to application of the test level .
T
24
IEC 60601 Test
Level
(>95% dip in
T
) for 0,5 cycle
T
(60% dip in
T
) for 5 cycles
T
(30% dip in
T
) for 25 cycles
T
(>95% dip in
T
) for 5 sec
T
Compliance Level
±6kV contact
±8kV air
±2kV for power
supply lines
±1kV for input/
output lines
±1kV differential
mode
±2kV common
mode
<5% U
(>95% dip
T
in U
) for 0,5 cycle
T
40% U
(60% dip
T
in U
) for 5 cycles
T
70% U
(30%
T
dip in U
) for 25
T
cycles
<5% U
(>95% dip
T
in U
) for 5 sec
T
3 A/m
Electromagnetic Environment
Guidance
Floors should be wood,
concrete or ceramic tile .
If floors are covered with
synthetic material, the relative
humidity should be at least
30% .
Mains power quality should be
that of a typical commercial or
hospital environment .
Mains power quality should be
that of a typical commercial or
hospital environment .
Mains power quality
should be that of a typical
commercial or hospital
environment . If the user of
the VenaFlow Elite requires
continued operation during
power mains interruptions,
it is recommended that the
VenaFlow Elite be powered
from an uninterrupted power
supply or a battery .
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment .