EMC GUIDANCE
Table 2
EN
Guidance and manufacturer's declaration – electromagnetic Immunity
Immunity Test
IEC 60601-1-2
DE
Test level
Electrostatic
±8 kV contact
FR
discharge (ESD)
±2 kV, ±4kV, ±8 kV, ±15 kV air
IEC 61000-4-2
ES
Electrical fast
Not applicable
IT
transient/burst
IEC 61000-4-4
Surge
Not applicable
IEC61000-4-5
Voltage dips, short
Not applicable
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
30 A/m
50Hz/60Hz
Power frequency
magnetic field
IEC 61000-4-8
Conduced RF
Not applicable
IEC61000-4-6
10 V/m
Radiated RF
80 MHz – 2,7 GHz
IEC61000-4-3
80 % AM at 1 kHz
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
Compliance level
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
Not applicable
Not applicable
Not applicable
30 A/m
50Hz/60Hz
Not applicable
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
34
Table 3
Guidance and manufacturer's declaration - electromagnetic Immunity
Radiated RF
Test
Band
Service
IEC61000-4-3
Frequency
(MHz)
(Test
(MHz)
specifications
for
385
380-390
TETRA
ENCLOSURE
400
PORT
IMMUNITY to
GMRS 460,
RF wireless
FRS 460
450
430-470
communica-
tions
710
704-787
LTE Band
equipment)
13,17
745
780
GSM
810
800-960
800/900,
TETRA 800,
870
iDEN 820,
CDMA 850,
930
LTE Band 5
GSM 1800;
1720
1700-
CDMA 1900;
1990
GSM 1900;
1845
DECT;
LTE Band 1,
3,
1970
4,25; UMTS
Bluetooth,
2450
2400-
WLAN,
2570
802.11
b/g/n, RFID
2450, LTE
Band 7
5240
5100-
WLAN
5800
802.11
5500
a/n
5785
EMC GUIDANCE
IMMUNITY
Modulation
Modulation
Distance (m)
TEST
(W)
LEVEL
(V/m)
1.8
0.3
27
Pulse
modulation b)
18Hz
FM c) ± 5kHz
2
0.3
28
deviation 1kHz
sine
Pulse
0.2
0.3
9
modulation b)
217Hz
Pulse
2
0.3
28
modulation b)
18Hz
Pulse
2
0.3
28
modulation b)
217Hz
Pulse
2
0.3
modulation
217 Hz
Pulse
0.2
0.3
modulation
217 Hz
35
EN
DE
FR
ES
IT
28
9