6.4
Evacuation of exhaled gases:
To safeguard the work environment and the staff, the exhaled gases must be conveyed
outside of the work environment.
Two systems can be used to discharge the gases externally:
-
Direct connection to the outside of the office
-
Connection to a surgical aspirator.
It is important to evaluate:
-
Whether the aspirator is suitable for the evacuation of the gases (N
-
The aspirator exhaust must be outdoors and ventilated.
Make sure it is compliant with regulations in force in the country of use.
Direct connection to the outside of the office
UNIT-MOUNTED MASTER FLUX
Screw the nozzles (Figure 3 Pos 8/9) onto the unit.
Remove the supplied valve (Figure 5 Pos 17).
Connect the mask circuit gas exhaust tube to the nozzle ( Figure 3 Pos 9).
Connect the gas evacuation spiral tube (Figure 9) to the nozzle
(Figure 3 Pos 8) on the unit.
Connect the end of the spiral tube to the outside and make sure there is a final filter.
WALL-MOUNTED MASTER FLUX
Connect the gas exhaust tube to the mask circuit to the nozzle ( Figure 9 Pos 36).
Connect the end of the spiral tube to the outside and make sure there is a final filter
(Figura 9 Pos 34).