Idiomas disponibles

• Error during setting up of USB drives
Error code
Display message
0x02000E01
Two USB keys
not supported
0x02000E21
...24
0x02000E25
...27
0x02000E28
...E29
USB connection
0x02000E2A
failed
0x02000E2B
0x02000E11-
0x02000E1B
• Error during printing and PDF generation
Error code
Display message
0x02001001
No USB drive
connected
0x02001013
No data to
print/save
0x02001011
...12
0x02001014
Failed to print/
save PDF
0x02001015
...17
Errors
Description
A max. of 1
A second USB drive was
USB drive
connected, but only one
supported
is supported
Internal USB interface
error
Error during
reading/opening of the
Error when
CSV file
installing USB
Printer has incorrect
drives to
device ID
the USB host
interface
Printer cannot be
initialized
The connected device is
neither a USB drive nor
a printer
Error when
installing USB
Analog to host
drives to the
(0x02000E21 -
USB OTG
0x02000E2B)
interface
Errors
Description
No USB drive
No USB drive
connected
recognized to save as
PDF file or printer to
print the measuring
results
No data in the
A USB memory medium
internal memory
or a printer is connected,
but no measured values
are available to output
Error during
The saving, creation or
creation/saving
writing of the PDF
of the PDF
document to a USB
document
drive has failed
Error while
Sending of commands
printing
to the printer failed
Remedy
Use only one USB
medium
Reconnect the USB
drive or restart the unit.
Format the USB drive if
necessary (FAT16/32)
Use a compatible printer
Disconnect the printer,
restart and reconnect it
Use compatible USB
drives
See above errors
Remedy
Connect the desired
USB drive (memory
medium / printer)
Perform the
measurement to fill the
memory with measured
values
Reconnect the USB
drive or restart the unit.
Format the USB drive if
necessary (FAT16/32)
Disconnect the printer,
restart and reconnect it.
Restart the ViscoClock
plus if necessary
(Disconnect the Power
supply from the DC
socket and then
reconnect it)