19
Connection:
BIOTRONIK cable
patient-side
Recommended connections
IS-1 connector
PK-67-S/L with PA-2
port
Possible connections
Alligator clips
PK-141
Alligator clips
PK-67-S/L with PA-4
Alligator clips
PK-67-S/L with PK-155 (2x)
Alligator clips
PK-82 (2x)
1
Only for 2-chamber pacing
Implanted lead with IS-1 connector
Polarity
Reocor D principally paces in bipolar mode, but it can be used
with bipolar or unipolar temporary pacing leads.
If unipolar leads are used, two leads must be connected for
each chamber.
Separating connections
Disconnect patient cables from the temporary pacing leads of
the patient or disengage the direct connection.
Separating Redel plug
Retract the retaining ring at the Redel plug and pull the
Redel plug off the Redel port.
Start Up
The operation of Reocor D is identical for all operating
modes. The operating steps should be carried out in the fol-
lowing order (the numbers in parentheses refer to Figure 1 on
page 11).
Insert battery.
Push protective cover down.
Prepare patient: Place the leads but do not connect them
to the pacemaker yet.
Prepare Reocor D:
Set the pacing rate with the Rate control dial (18).
Set the AV delay with the AV delay control dial (15)
Set the pacing amplitudes for atrium
the Atrium Ampl. (10)
dials.
Connection
Reocor D
device-side
connection
Redel plug
Redel adapter
Redel plug
Redel adapter
Redel plug
Redel adapter
Redel plug
Redel adapter
2-mm plug
2-mm connector port
1
and ventricle with
1
and Ventricle Ampl. (6) control
1
.