Electromagnetic Compatibility Information
68
Table 3
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The custo-
mer or the user of the device should assure that it is used in such an environment.
IMMUNITY test
Conducted
RF IEC
61000-4-6
Radiated RF
IEC 61000-
4-3
www.spengler.fr
IEC 60601test
Compliance level
level
3V for 0.15-
3V for 0.15-
80MHz; 6V in
80MHz; 6V in
ISM and amate
ISM and amate
-ur radio bands
-ur radio bands
between0.15-
between0.15-
80MHz
80MHz
385MHz, 27V/m
385MHz, 27V/m
450MHz, 28V/m
450MHz, 28V/m
710MHz, 745
710MHz, 745
MHZ,780MHz
MHZ,780MHz 9V/m
9V/m
810MHz, 870
810MHz, 870
MHZ,930MHz
MHZ,930MHz
28V/m
28V/m
1720MHz, 1845
1720MHz, 1845
MHZ, 1970MHz
MHZ, 1970MHz
28V/m
28V/m
2450MHz, 28V/m
2450MHz, 28V/m
5240MHz, 5500
5240MHz, 5500
MHZ, 5785MHz
MHZ, 5785MHz
9V/m
9V/m
Electromagnetic environment-gui-
dance
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the device, inclu-
ding cables, than the recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Recommended seperation distance
80 MHz to 800 MHz
800 MHz to 2,7 Ghz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transm-
itter manufacturer and d is the
recommended separation distance in
metres (m).
Field strengths from fixed RF transmit-
ters, as determined by an zelectro-
magnetic site survey, a should be less
than the compliance level in each
frequency range.
Interference may occur in the vicinity
of equipment marked with the fol-