5.3
ORP Sensor Calibration
Only required for TRUSENSE model.
NOTE: The ORP sensor wire should remain connected during calibration.
1.
Turn off pump and filtration system at the breaker. Automation
System needs to be in auto, NOT service mode.
2.
Close the valves if needed.
3.
Disconnect the Red ORP sensor BNC connector from the RED
labeled right side connector of the electronic circuit board
4.
Fill the storage tube to the halfway point with ORP 470 buffer
solution, see Figure 11.
Figure 11.
Fill storage tube halfway
5.
Loosen the right compression fitting and lift ORP sensor carefully
from plastic housing, DO NOT touch the glass tube.
6.
Rinse the tip of the ORP sensor with clean water.
7.
Shake ORP sensor to remove residual water. Do NOT touch the gold
tip at the end of the sensor.
8.
Connect the Red ORP sensor BNC connector to the RED labeled right
side connector of the electronic circuit board.
9.
Submerge sensor into the ORP 470 mV buffer solution inside the
storage tube. Start calibration using and following the iAquaLink
or OneTouch controller (AquaLink
SYSTEM SETUP > TRUSENSE, see Figure 12.
Jandy
TruSense Water Chemistry Analyzer | Owner's Manual
®
RS only) go to: MENU / HELP >
®