TruSense Water Chemistry Analyzer | Owner's Manual
®
Figure 12.
Submerge sensor & start calibration
10. Remove the ORP sensor from storage tube and rinse the tip of the
sensor with clean water.
11. Once calibration is complete, carefully insert sensor back into Jandy
TruSense and hand tighten the compression fitting.
Section 6.
Sensors must be regularly visually checked and cleaned if there is
visible buildup.
1.
Turn off pump and filtration system at the breaker. Automation System
needs to be in auto, NOT service mode.
2.
Close the valves if needed.
3.
Disconnect the sensor BNC connector from the electronic circuit
board.
4.
Loosen the compression fitting and lift the sensor carefully from
plastic housing.
5.
Rinse the sensor with clean water. Shake off excess water.
6.
Gently brush the junctions, glass bulb, and metal strip using a new or
clean toothbrush for 1 minute.
7.
Prepare a diluted solution of muriatic acid (10 drops in 2 fl. oz of tap
water).
8.
Wash the sensor in the diluted acid solution for 2 minutes.
Sensor Cleaning
ENGLISH
Calibrate pH
Page 19