Miscellaneous; Cleaning And Maintenance - Medisana BTS Instrucciones De Manejo

Tonificador muscular electronico
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 63
GB
4 Application / 5 Miscellaneous
4.5
Errors and repair

5 Miscellaneous

5.1
Cleaning and
maintenance
26
The device doesn't turn on after the ON/OFF switch is pressed.
The batteries are not inserted correctly. Insert the batteries correctly.
The batteries are empty. Replace the batteries.
The muscle only contracts a little although the intensity is set high.
The battery is almost empty. Replace the battery.
The electrodes are worn out / don't stick properly. Order new electrodes
from MEDISANA.
The electrodes are placed too close together. Both electrodes from a single
patient lead are to be placed as far away as possible from each other on
one muscle.
Unpleasant feeling where the electrodes are placed.
The electrodes are poorly placed. Check the placing of the electrodes
according to the instructions for use.
The electrodes are worn out / don't stick properly. Order new electrodes
from MEDISANA.
A pair of electrodes doesn't emit any stimulus.
The intensity level of the exit channel is set at zero (0). Increase the
intensity.
The cables are not inserted correctly. Check whether all the cables are ins
erted firmly in the connector plugs.
One cable is faulty. Insert a different cable in the connector socket. If it
works the first cable is faulty and should be exchanged.
One plug is faulty. Insert the cable in a different connector socket. If it
works, the plug is faulty. Bring the device to your specialist dealer.
• Remove the batteries before cleaning the unit.
• Never use strong detergents or hard brushes.
• Clean the unit using a soft cloth lightly moistened with a mild soapy
solution. Never use abrasive cleaning agents or alcohol. Do not allow any
water to get into the unit. Do not use the unit again until it is completely dry.
• Do not expose the device to direct sunlight or to high temperatures and
protect it from dust, dirt and damp.
• Store the electrodes between applications on the protective film in the re-
sealable storage pouch in a cool dry place.
• The effectiveness of the electrodes gradually subsides after 20–30 applica-
tions due to skin deposits in the gel. To maintain the adhesiveness, hold the
electrodes under cold running water for a few seconds and then shake them
dry. Before using the electrodes again, leave them on the protective film for
at least 8 hours.
• Remove the batteries from the unit if you do not plan on using it again
anytime in the near future. Otherwise, the batteries may start leaking.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

87302

Tabla de contenido