Hygiene treatment and disposal
4.2 Disinfecting the exhalation system
1. Clean individual parts before disinfection
exhalation system", page
2. Immerse mask parts in Sekusept
concentration of 2% for 15 minutes.
3. Wash immersed individual parts with a soft brush for at least 3
minutes.
4. Pay attention to all creases and cavities.
5. Ensure that there are no air bubbles on the individual parts.
6. Rinse individual parts after disinfection for at least 1 minute with
water (at least drinking water quality, sterile distilled water or
fully demineralized water).
7. Allow the individual parts to air-dry.
8. Assemble the exhalation system
exhalation system", page
1
A bactericidal, levurocidal, tuberculocidal, mycobactericidal,
sporicidal, fungicidal (clean cond.) and virucidal effect is achieved at
this concentration and time to take effect.
2
Recommendations of KRINKO BfArM [Kommission für
Krankenhaushygiene und Infektionsprävention - Commission for
Hygiene and Infection Prevention at the Robert Koch Institute/
Bundesinstitut für Arzneimittel und Medizinprodukte - Federal
Institute for Drugs and Medical Devices] (pages 1252/1254).
26).
®
Aktiv solution at a
1
2
(see "3.1 Assembling the
23).
(see "4.1 Cleaning the
|
EN
27