NEBULIzER CLEANING
All parts of the nebulizer, except tubing, should be cleaned according to the following
instructions. Your physician and/or DeVilbiss provider may specify a certain cleaning
procedure. If so, follow their recommendations.
NOTE– Before initial operation, your nebulizer should be cleaned following cleaning
instructions, or as recommended by your physician or DeVilbiss provider.
WARNING
To prevent possible risk of infection from contaminated medication, cleaning of the
nebulizer is recommended after each aerosol treatment. Disinfecting is recommended
once a day.
Clean After Every Use:
1.
With power switch in the "Off" (O) position,
unplug power cord from wall outlet.
2.
Disconnect tubing from the air-inlet connector and set aside.
3.
Disassemble mouthpiece or mask from cap. Open nebulizer by turning cap
counterclockwise and removing baffle.
4.
Wash all items, except tubing, in a warm water/dishwashing detergent solution. Rinse
under warm tap water for 30 seconds to remove detergent residue. Allow to air dry.
disinfect daily:
1.
Using a clean container or bowl, soak items in three parts warm water
(131ºF-149ºF/55ºC-65ºC) to one part white vinegar (>=5% acetic acid concentration)
for 30 minutes (Figure 9) OR use a medical bacterial-germicidal disinfectant available
through your provider. For reusable nebulizer only, clean daily in dishwasher using top
shelf. Be sure to follow manufacturer's instructions carefully.
2.
With clean hands, remove items from disinfectant solution, rinse under warm tap water,
and air dry on a clean paper towel. Store in a zip-lock bag.
NOTE– Do not towel dry nebulizer parts; this could cause contamination.
CAUTION– The reusable nebulizer is dishwasher safe, but do not place any disposable
nebulizer parts in an automatic dishwasher; doing so could cause damage.
WARNING
To prevent possible risk of infection from contaminated cleaning solutions, always
prepare fresh solution for each cleaning cycle and discard solution after each use.
3.
Keep the outer surface of the tubing dust-free by wiping regularly. Nebulizer tubing
does not have to be washed because only filtered air passes through it.
NOTE– The reusable nebulizer will last up to one year. Proper cleaning will help extend the
life of the nebulizer. We recommend that an extra nebulizer be kept on hand at all times. The
reusable nebulizer is dishwasher safe and may be cleaned and reused for up to one year.
COMPRESSOR CLEANING
1.
With power switch in the "Off" (O) position, unplug power cord from wall outlet.
2.
Wipe outside of the compressor cabinet with a clean damp cloth every few days to
keep dust-free.
dANGER
Do not submerge in water; doing so will result in compressor damage.
FILTER CHANGE
1.
F ilter should be changed every 6 months or sooner if filter turns completely gray in
color.
2.
R emove filter cap by grasping it firmly and pulling out of the unit (Figure A).
3.
R emove dirty filter with fingers and discard.
4.
R eplace with a new DeVilbiss filter (part #5650D-602). Additional filters should be
purchased from your DeVilbiss provider.
5.
P ush filter cap with new filter into position.
CAUTION– Reusing filter or substituting any other material such as cotton for a DeVilbiss
filter will result in compressor damage.
NOTE– An extra filter is packaged with the 4650 series units.
SE-650
Figure 4
Figure 5
Figure 6
Figure 7
Figure 8
Figure 9
EN - 5