General warnings
• Only use the Flocare Infinity enteral feeding pump for delivery of enteral feeding solutions or hydration fluids and
liquids. Do not use for intravenous administration of infusions.
• Before using or setting up a feeding regimen, check the integrity of the pump. Do not use the pump if it is damaged.
If a technical failure occurs or if the pump is dropped, the pump should be checked by a qualified technician.
• A leaking enteral feeding system might compromise the pumps' ability to detect an occlusion. Ensure the enteral
feeding system is installed and connected properly. During a feeding therapy, ensure there is no leakage observed
within the system.
In case of any leakage, tighten the connectors and closure caps and verify whether leakage stopped. In case the
leakage still occurs, replace the Flocare Infinity giving set.
• In the event of malfunction of the pump or changes in its performance, please consult your local contact Sales
Representative or care giver for a pump service inspection or replacement as required.
• Plug the pump charger into the mains to charge the battery before commencing enteral feeding. If the pump has
been stored for any period of time, it should be plugged into the mains to recharge the battery before commencing
enteral feeding. The battery will be completely charged after approximately 6 hours. The battery capacity is an
approximation. If you are unsure that enough capacity remains for your intended use, recharge it.
• Use only approved cleaning agents, using non-approved cleaning agents might harm the pump.
• To avoid electrical shock, switch the pump off and remove the plug from the mains before cleaning the pump.
• Make sure the Infinity AC power adapter is completely dry before plugging into the mains.
• Do not use the Infinity alarm systems to trigger actions related to secondary (electrical) medical devices (e.g. a
volumetric or syringe pump).
• Do not use pump functions for any other purpose than described in this manual.
E.g. FILL SET function to overcome an occlusion or the AIR in giving set alarm to indicate end of therapy.
Precautions
• The Flocare Infinity enteral feeding pump can be used for adult and paediatric patients. The health care professional
is responsible to determine the therapy setting and clinical needs, as well as the appropriate surveillance scheme
and its frequency. If the required surveillance scheme cannot be guaranteed by the care giver, then it is to be
discussed with the healthcare professional, who can advise the patient an alternative therapy solution.
• For volume and nutrient sensitive patients (e.g. children with metabolic conditions using overnight and/or
unmonitored therapy) where a constant nutrition delivery is vital and/or connected with consequential therapies (e.g.
an insulin delivery pump), additional surveillance is recommended to be put in place to verify good functioning of
the enteral feeding system and programmed therapy.
• The sound level of the Flocare Infinity pumps audible alarm is to be set in 'HIGH' if the pump is operated in a noisy
environment or the healthcare provider is not nearby the pump, such as at night, to ensure the alarm is readily
noticed when an alarm is activated.
• It is recommended to use ready to feed tube nutrition as this is developed for the intended delivery specifications
of the pump.
• Use only feeding solutions prescribed by the responsible clinician, registered dietician, registered nurse or other
licensed practitioner.
• Before commencing any enteral feeding therapy, proper consideration is to be given to the feed appropriateness
for tube feeding (thickness, homogeneity, selection of nasogastric tube Ch size, etc.), as to avoid occlusion or any
other unexpected pump system behaviour. The condition and appropriateness of the nutrition can be addressed
with the healthcare professional who knows the clinical situation and needs of the patient and can best guide the
patient in the selection of appropriate feed.
403259675_r0.indd 19
403259675_r0.indd 19
EN
WARNINGS AND PRECAUTIONS
-19-
13/04/23 3:34 AM
13/04/23 3:34 AM