damage, cracking, colour changes, etc. to the medical device during and after the application. In case of doubt,
contact your local (Nutricia) sales representative.
• Avoid harsh or agressive cleaners/ disinfectants such as acetone, benzene or iodine.
• Do not sterilize the pump by irradiation, EtO, steam or in an autoclave.
• Always unplug the pump prior to cleaning to avoid electric shock hazard.
• On a regular basis thoroughly clean all surfaces of the pump (including the sensors and rotor) with warm soapy
water, a 5% bleach solution in water, a multipurpose disinfectant cleaner or a soft cloth.
• The Flocare
Infinity
®
pump!
• Always maintain the rollers on the rotor in a clean state to ensure they spin freely.
• The charger normally does not require cleaning. When desired, a dry or slightly damp cloth may be used
to clean the outside surface of the charger. Make sure the charger is disconnected from the wall outlet.
• Pump disposal: At the end of their service lifetime , the pump and its electrical accessories (AC Adapter Charger
and Data Download Accessory Cable), should be disposed of according to local standards and regulations
governing the disposal of electronic waste (e-waste). Other accessories can be disposed of or recycled as standard
non-hazardous waste.
In case any of the problems listed on the problem solver chart occur, the pump delivers an audible and visual alarm
and stops working. The back light of the LCD automatically switches on.
Exception to this is the low battery alarm "BATT", in this situation the pump continues working.
Action in the event of an alarm:
Check the type of alarm displayed by the large display characters.
• Press the "START/STOP" key to stop the audible and visual alarm.
• Correct the cause of the alarm as described in the table.
• Start the pump again by pressing "START/STOP".
The problem solver chart gives a clear explanation of the alarms.