Descargar Imprimir esta página

Fisher Scientific Fisherbrand Isotemp Instalación Y Operación página 16

Publicidad

Idiomas disponibles

Idiomas disponibles

Max and Min Temperature
Memorization
How to See the Min Temperature
1. Press and release the
2. The "Lo" message will be displayed followed by the
minimum temperature recorded.
3. By pressing the
normal display will be restored.
How to See the Max Temperature
1. Press and release the
2. The "Hi" message will be displayed followed by the
maximum temperature recorded.
3. By pressing the
key again or by waiting 5s the
normal display will be restored.
How to Reset the Max and Min
Temperature Recorded
1. Hold press the SET key for more than 3s, while the
max. or min temperature is displayed. (rSt message will
be displayed).
2. To confi rm the operation the "rSt" message starts
blinking and the normal temperature will be displayed.
16
key.
key again or by waiting 5s the
key.
Main Functions
How to See the Setpoint
1. Push and immediately release the
SET key: the display will show the
Set point value.
2. Push and immediately release the
SET key or wait for 5 sec to display
the probe value again.
How to Change the Setpoint
1. Push the SET key more than 2 sec to change the Set
point value.
2. The value of the set point will be displayed and the
LED will start blinking.
3. To change the Set value push the
10 sec.
4. To save the new set point value, push the SET key
again or wait for 10 sec.
How to Lock the Keyboard
1. Hold the
more than 3s.
2. The "POF" message will be
displayed and the keyboard will
be locked. At this point, it will be
possible only to see the set point or
the Max or Min temperature stored.
3. If a key is pressed more than 3s the
"POF" message will be displayed.
+
arrows within
and
keys for

Publicidad

loading