• ACTUATORS (Electric models only): The actuators should require no maintenance other
than checking for correct operation and listening for unusual noise. Ensure all fixing points
are secure. Confirm the anti-crush mechanism is operational (lifting actuator only). If in
doubt, DO NOT use and have the unit checked by a qualified person or authorised Oxford
Service Provider.
• EMERGENCY RAISE/LOWER AND EMERGENCY DESCENT: Check both the redundant
raise/lower controls and the mechanical emergency descent function (if fitted) using the
maximum safe working load applied to the end of the boom. Please note that the practice
of using a spring balance will not correctly check the function of the emergency descent.
• HYDRAULIC UNIT: The hydraulic unit should require no maintenance other than checking
for correct operation in both directions and leakage of hydraulic fluid. Additionally, check
the ram for trapped air/sponginess.
• BATTERIES (Electric models only): The batteries are located in the Power Pack and
should not require maintenance other than the regular charging as detailed in the charging
instructions.
• CLEANING: Clean with ordinary soap and water and/or any hard surface disinfectant.
Harsh chemical cleaners or abrasives should be avoided as these may damage the
surface finish of the lift. Avoid wetting any of the electrical parts.
• SLINGS: Check for wear and fraying. DO NOT use any sling that is damaged, or if the
sling shows signs of wear.
• LOAD TEST: The load test should be carried out in accordance with the
manufacturer's test procedures. It is strongly recommended the testing is carried out by an
authorised service dealer.
• CERTIFICATION: An authorised service dealer will issue a test certificate after satisfactory
completion of the load test. This certificate will be valid for six months.