LOW CURRENT
It shows that the test current is not enough to carry out the reading. It
appears at the beginning of each test and it keeps on being there up to
the operator rotates the Test current control
for the test current to be higher than 10 % of the nominal current in the
range. The inappropriate connection of the cables may cause a difficult
circulation of test current. If this message keeps on being displayed,
please check that the current cables are connected appropriately.
OVERRANGE
It indicates that the measured resistance is higher than the maximum
value readable in the selected range.
MEM
It appears at the beginning of the second line of the display, showing that
the value is the one retained in the display when pressing the Hold
button.
OVERHEATING
It indicates that some part of the equipment has achieved the critical
temperature. Thus the system will cut the high current generation. This
message appears only when the 200 A range is selected. It behaves as a
thermal protection.
BAT
It shows that the battery is quite discharged. It is necessary to recharge
the battery.
clockwise, as necessary