Optima Prone / English
6. Cleaning
It is important to follow the cleaning procedures to avoid cross contamination. Be sure to clean the
surface in a dry and dust free environment. Wipe down the pump unit with a damp cloth pre-soaked
with a mild detergent. Avoid contact with dust and proximity to dusty areas. Make sure that any
cleaning agents you use will not harm or corrode the plastic casing on the pump unit. If your doctor or
medical facilities have other special cleaning instruction, please follow the professional instruction.
CAUTION: Do not immerse or soak pump unit/position sensor.
WARNING: Do not remove the casing of the pump to avoid electrical shock. All disassembly
or repair should be done by authorized technicians.
CAUTION: The pump does not need oil lubrication; please do not dissemble the system.
Cover Material:
Stretch
Cover Material:
Nylon/PU
Wipe-down the mattress unit with a damp cloth pre-soaked with warm water containing a mild
detergent, or chlorine bleach followed by an approved intermediate level disinfectant. Also the mattress
top cover can be completely removed for laundry with water temperature indicated on the laundry
label; however, it is recommended that the user still check with local policy to determine the
time/temperature ratio required achieving thermal disinfection. The cover may also be cleaned using
sodium hypochlorite diluted in water. After cleaning, please avoid dust and proximity to dusty areas and
all parts should be air dried thoroughly before use.
CAUTION: Do not use phenolic based products for cleaning.
CAUTION: Keep mattresses out of direct exposure to sunlight, and make sure to dry the
mattress after cleaning.
CAUTION: Make sure to execute the cleaning procedure every time before a new patient lying
on the air mattress.
21